Challenges With In Situ X-Ray Tomography Mechanical Testing on Small Scale Nuclear Materials |
| Author(s) |
Alex G. Arzoumanidis, Matthew E. Struve |
| On-Site Speaker (Planned) |
Matthew E. Struve |
| Abstract Scope |
Small scale mechanical testing on irradiated specimens provides more physical samples when material quantities are limited, with the added safety benefit of reduced radioactive exposure. A high force, dual column, rotating load train universal test system is presented, including fixturing for miniature irradiated dogbone samples. Bearing runout is less than 1 micron and a four degree of freedom fixture aligns the two rotating axes. The open nature of the dual column frame enables laser heating for extreme temperature testing. |