| Presentation Title |
Optimization of the Calibration Curve for WDXRF Analysis of Bayer Process Residues through Expansion of the Reference Matrix |
| Author(s) |
Carlos Silva Oliveira, Tracy Fonseca, Antonio Paulino Neto, Jordania Vasconcelos Abreu, Joao Renan |
| On-Site Speaker (Planned) |
Carlos Silva Oliveira |
| Abstract Scope |
The accuracy of wavelength-dispersive X-ray fluorescence (WDXRF) analysis strongly depends on the robustness of the calibration matrix. This work presents the optimization of a WDXRF calibration for Bayer process residues through the expansion of the reference matrix using certified reference materials and synthetic standards. The calibration range was extended for Na2O, V2O5, P2O5, and MnO, eliminating analytical extrapolation and improving calibration performance. The optimized calibration was evaluated using correlation coefficients, root mean square error (RMSE), and statistical comparison with the original model. A substantial reduction in prediction error was achieved, particularly for Na2O, whose RMSE decreased from 0.90 wt.% to 0.057 wt.%. The expanded calibration improves analytical reliability for routine quality control and process monitoring in alumina refineries, providing a practical methodology for enhancing WDXRF calibration applied to Bayer process residues. |